Modifications to Willmore's method for calibrating seismographs using a maxwell bridge
Auteur(s): |
R. E. White
|
---|---|
Médium: | article de revue |
Langue(s): | anglais |
Publié dans: | Bulletin of the Seismological Society of America, 1 décembre 1970, n. 6, v. 60 |
Page(s): | 2015-2022 |
DOI: | 10.1785/bssa0600062015 |
Abstrait: |
The substitution of an oscilloscope as detector into a Willmore bridge circuit allows a simple, accurate, and direct method of calibration for seismometer systems. An example of the method is given and its modification for a seismometer having appreciable inductance is discussed. It is also shown that transient methods do not allow an accurate determination of high-frequency response. |