Modifications to Willmore's method for calibrating seismographs using a maxwell bridge
Autor(en): |
R. E. White
|
---|---|
Medium: | Fachartikel |
Sprache(n): | Englisch |
Veröffentlicht in: | Bulletin of the Seismological Society of America, 1 Dezember 1970, n. 6, v. 60 |
Seite(n): | 2015-2022 |
DOI: | 10.1785/bssa0600062015 |
Abstrakt: |
The substitution of an oscilloscope as detector into a Willmore bridge circuit allows a simple, accurate, and direct method of calibration for seismometer systems. An example of the method is given and its modification for a seismometer having appreciable inductance is discussed. It is also shown that transient methods do not allow an accurate determination of high-frequency response. |