Evaluation of the effects of leakage current in the calibration of standard resistors between two dc resistance bridges and a digital sampling impedance bridge

Auteur(s):

Médium: article de revue
Langue(s): anglais
Publié dans: Journal of Physics: Conference Series, , n. 1, v. 2606
Page(s): 012032
DOI: 10.1088/1742-6596/2606/1/012032
Abstrait:

This article discusses calibrations results of standard resistors performed with two commercially available equipment and a digital sampling impedance bridge developed at Inmetro. One of the commercial bridges uses the current comparison method and the other the potentiometric method with a binary voltage divider. Calibration was performed with standard resistors of 10 kΩ and 100 kΩ with significant leakage current. This study showed divergent results between each methodology. A specific methodology was used to estimate the leakage resistance of these resistors. This study contributes to establish criteria for correction and/or inclusion of uncertainty components in resistors calibrated in these methodologies.

Structurae ne peut pas vous offrir cette publication en texte intégral pour l'instant. Le texte intégral est accessible chez l'éditeur. DOI: 10.1088/1742-6596/2606/1/012032.