Evaluation of the effects of leakage current in the calibration of standard resistors between two dc resistance bridges and a digital sampling impedance bridge
Auteur(s): |
Márcio Cândido da Silva
Gean Marcos Geronymo |
---|---|
Médium: | article de revue |
Langue(s): | anglais |
Publié dans: | Journal of Physics: Conference Series, 1 octobre 2023, n. 1, v. 2606 |
Page(s): | 012032 |
DOI: | 10.1088/1742-6596/2606/1/012032 |
Abstrait: |
This article discusses calibrations results of standard resistors performed with two commercially available equipment and a digital sampling impedance bridge developed at Inmetro. One of the commercial bridges uses the current comparison method and the other the potentiometric method with a binary voltage divider. Calibration was performed with standard resistors of 10 kΩ and 100 kΩ with significant leakage current. This study showed divergent results between each methodology. A specific methodology was used to estimate the leakage resistance of these resistors. This study contributes to establish criteria for correction and/or inclusion of uncertainty components in resistors calibrated in these methodologies. |
- Informations
sur cette fiche - Reference-ID
10777723 - Publié(e) le:
12.05.2024 - Modifié(e) le:
12.05.2024